Commissioning and benchmarking a 3D dosimetry system for clinical use

94

Citations

38

References

2011

Year

Abstract

This work presents the first comprehensive benchmarking of a 3D dosimetry system for clinical use. The DLOS/PRESAGE benchmark tests show consistently good agreement to simple known distributions. The system produces accurate isotropic 2 mm dose data over clinical volumes (e.g., 16 cm diameter phantoms, 12 cm height), in under 15 min. It represents a uniquely useful and versatile new tool for commissioning and verification of complex therapy treatments.