<b>Late Stent-Graft Migration Secondary to Separation of the Uncovered Segment From the Main Body of a Zenith Endoluminal Graft</b>

Karim Ghanim, Bibombe P. Mwipatayi, Manzoor Abbas, Kishore Sieunarine

Journal of Endovascular Therapy · 2006 · 12 citations · 7 references

Concepts

Abstract

PURPOSE: To report a case of type I endoleak secondary to complete disruption of the sutures uniting the uncovered and covered segments of a bifurcated Zenith endoluminal graft, causing displacement and distal migration of the graft main body. CASE REPORT: A 76-year-old man had successful exclusion of an abdominal aortic aneurysm with a Zenith endoluminal graft in 1999. He continued to do well until the 4-year surveillance imaging [computed tomography (CT) and plain abdominal radiography] showed device migration and proximal endoleak, with consequent expansion of the aneurysm. A proximal extension stent-graft was inserted with good seal. The 1-month follow-up CT angiogram showed reduced aneurysm size and no evidence of any leak. CONCLUSION: This case shows that the failure of an endoluminal graft occurs at weak points in the construction of the graft, reinforcing the need for long-term surveillance. If detected promptly, such events can often be treated by another endovascular procedure.

References

7