Estimated Radiation Dose Reduction Using Adaptive Statistical Iterative Reconstruction in Coronary CT Angiography: The ERASIR Study
300
Citations
38
References
2010
Year
There was no difference between groups in the use of prospective gating, tube voltage, or scan length. The examinations performed using ASIR had a lower median tube current than those obtained using FBP (median [interquartile range], 450 mA [350-600] vs 650 mA [531-750], respectively; p < 0.001). There was a 44% reduction in the median radiation dose between the FBP and ASIR cohorts (4.1 mSv [2.3-5.2] vs 2.3 mSv [1.9-3.5]; p < 0.001). After adjustment for scan settings, ASIR was associated with a 27% reduction in radiation dose compared with FBP (95% CI, 21-32%; p < 0.001). Despite the reduced current, ASIR was not associated with a difference in adjusted signal, noise, or signal-to-noise ratio (p = not significant). No differences existed between FBP and ASIR for interpretability per coronary artery (98.5% vs 99.3%, respectively; p = 0.12) or per patient (96.1% vs 97.1%, p = 0.65). CONCLUSION. ASIR enabled reduced tube current and lower radiation dose in comparison with FBP, with preserved signal, noise, and study interpretability, in a large multicenter cohort. ASIR represents a new technique to reduce radiation dose in coronary CTA studies.