Publication | Closed Access
Fretting and Corrosion in Modular-Neck Total Hip Arthroplasty Femoral Stems
112
Citations
21
References
2014
Year
The ABG II dual modular hip system is associated with a high rate of early failure secondary to fretting and corrosion at the femoral neck-stem taper. The component has subsequently been recalled and is no longer in use. Surgeons using modular hip systems with a titanium stem and cobalt-chromium neck should be vigilant about annual follow-up with radiographs, and use of MRIs as indicated.