Communications in Soil Science and Plant Analysis · 1984 · 106 citations · 25 references
Natural SciencesChemistryPlant SciencesAnalytical ChemistryElemental AnalysisMethod ValidationAnalytical MeasurementsElemental Composition AnalysisInstrumental AnalysisPlant Tissue AnalysisSample PreparationTrace Element AnalysisAssay Data QualityAssay Data Quality ControlAssay Quality ControlCalibration ProceduresInstrument CalibrationMultielement DeterminationPrecisionQuantitative Analysis Of Elements
ICP‑AES is increasingly used for plant analysis, but inter‑laboratory quality control is lacking and requires standardized preparation and calibration to ensure reliable multi‑element data. The laboratory’s quality control initiatives are presented. A survey of 12 service laboratories and an inter‑laboratory comparison of 8 plant samples sent to 5 ICP‑AES labs with varying preparation and calibration protocols were conducted. Precision varied across elements, with P, Mg, Mn, Ca, and K showing the best reproducibility, while Al, Zn, Fe, Na, Cu, and B were most imprecise, likely due to differing preparation and calibration practices. Keywords: ICP spectroscopy, plant tissue analysis, quality assurance.
Abstract A survey of 12 service laboratories using ICP‐AES for routine plant analysis reveals a variety of sample preparation and instrument calibration procedures in use. With the continued growth of ICP applications in plant analysis laboratories, a need exists for inter‐laboratory quality control. The assurance of data quality among laboratories cannot be expected unless a standardization of preparation methods and calibration is accomplished. In the selection of a method for multi‐element determination on the same sample solution, the completeness of elemental release and sensitivity for some elements may have to be compromised. In addition to the above survey, 8 plant samples were sent to 5 ICP laboratories that handle large volumes of plant samples with each using different preparation and calibration methods. The results for 11 elements show that the best precision among laboratories was for P, Mg, Mn, Ca and K, and the poorest precision for Al, Zn, Fe, Na, Cu and B. The imprecision noted for some elements was presumed to be caused by the diversity of preparation and calibration methods. Quality control efforts taken by this laboratory will also be given. Key words: ICP spectroscopyplant tissue analysisquality assurance
25
Official Methods of Analysis of
Analytical Chemistry · 1980 · 59.6K citations
Management, Educational Assessment, Information Retrieval +5
Diagnostic Criteria for Plants and Soils
BioScience · 1966 · 964 citations
Soil Testing and Plant Analysis
P. W. Arnold · Geoderma · 1974 · 918 citations
W. J. Dixon · Biometrics · 1953 · 868 citations