Muscle & Nerve · 1992 · 49 citations · 11 references
Life SciencesBiomedical EngineeringNeuroscienceMotor NeuroscienceNeural EngineeringNeurophysiologyMotor Control PhysiologyMotor Neuron PhysiologyMotor PathwaysMotor SystemsMuscle Signal ProcessingNeurostimulationActivationBlockingElectrical Muscle StimulationJitterMotor Endplate PhysiologyNeural Stimulation ThresholdStimulation Site Localization
Electrical microstimulation of motor axons in conjunction with single fiber EMG (SFEMG) is increasingly used to measure the jitter of the motor endplates. This study examines the jitter of the stimulation site on the axon when stimulus strength is at threshold. In the absence of spurious blocking, this was found to be 5 microseconds on the average. With intermittent blocking, however, a mean additional jitter of 40 microseconds was obtained. The latter is considered to result from changing propagation velocity in the muscle fiber due to irregularity of activation rate. In clinical jitter studies, inadvertent threshold stimulation can result in significant error when associated with intermittent blocking.
11
Axonal stimulation for end-plate jitter studies.
J. V. Trontelj, Marjan Mihelin, J M Fernandez et al. · Journal of Neurology Neurosurgery & Psychiatry · 1986 · 118 citations · Full text