Metrologia · 1972 · 23 citations · 13 references
EngineeringElectrical EngineeringHeat TransferRadiation MeasurementSensorsInfrared InstrumentationRadiation InstrumentationTemperature MeasurementTemperature SensorsBlackbody RadiationEmissivityEmissivity EstimationExperimental TestingLampsRadiatorsSurface EmissivitySystematic Uncertainties
A description is presented of the establishment of the IPTS68 from 1064.43 °C to 2200 °C on vacuum and gas-filled tungsten blackbody lamps using the NSL photoelectric pyrometer. The apparatus, experimental technique and computational methods, including a method for calculating the effective emissivity of the walls of blackbody cavity radiators, are described and the components of all significant uncertainties are stated. The total uncertainties (99% confidence limits) obtained are ±0.10 °C at 1064.43 °C and ±0.25 °C at 1627 °C for vacuum lamps, and ±0.4 °C at 1627 °C and ±1.6 °C at 2200 °C for the gas-filled lamp.
13
A Precise Measurement of the Stefan-Boltzmann Constant
W. R. Blevin, William O. Brown · Metrologia · 1971 · 103 citations
International Practical Temperature Scale of 1968