The Realization of the IPTS68 above 1064.43 °C Using the NSL Photoelectric Pyrometer

T P Jones, J Tapping

Metrologia · 1972 · 23 citations · 13 references

Concepts

Abstract

A description is presented of the establishment of the IPTS68 from 1064.43 °C to 2200 °C on vacuum and gas-filled tungsten blackbody lamps using the NSL photoelectric pyrometer. The apparatus, experimental technique and computational methods, including a method for calculating the effective emissivity of the walls of blackbody cavity radiators, are described and the components of all significant uncertainties are stated. The total uncertainties (99% confidence limits) obtained are ±0.10 °C at 1064.43 °C and ±0.25 °C at 1627 °C for vacuum lamps, and ±0.4 °C at 1627 °C and ±1.6 °C at 2200 °C for the gas-filled lamp.

References