We demonstrated the feasibility of using a large number of PSDs in a new 2D-PSDA for the dosimetry of MV energy photon beams in radiation therapy. The excellent precision, accuracy, and low angular dependence of the device indicate that such a prototype could potentially be used as a high-accuracy quality assurance tool for IMRT and arc therapy patient plan verification. The homogeneity and water-equivalence of the prototype we built suggest that this technology could be extended to multiple detection planes by arranging the fibers into more complex orientations, opening the possibility for 3D dosimetry with PSDs.