Concepedia

Publication | Open Access

Long-Term Evolution of the Electrical Stimulation Levels for Cochlear Implant Patients

36

Citations

18

References

2012

Year

Abstract

We have studied the long-term evolution of the stimulation levels in cochlear implant patients. Our results show the importance of systematic measurements of the electrode impedances before the revision of the programming map. This report also highlights that the evolution of the programming maps is an important factor to be considered in order to determine an adequate calendar fitting of the cochlear implant processor.