Testing the Honeywell Durafet® for seawater pH applications

Todd R. Martz, James G. Connery, Kenneth S. Johnson

Limnology and Oceanography Methods · 2010 · 301 citations · 19 references

DOIFull text

Open access

Concepts

TL;DR

We present a theoretical overview of sensor response using both internal and secondary reference electrodes. The sensor was evaluated in temperature‑controlled calibration vessels, the MBARI test tank, shipboard mapping, and a surface mooring, with many tests employing a secondary reference electrode in addition to the internal reference, and a theoretical overview of sensor response was provided. The Honeywell Durafet pH sensor, tested in seawater at 1 atm, shows short‑term precision of ±0.0005 pH over hours, stability better than 0.005 pH over weeks to months, and performance sufficient for many low‑pressure biogeochemical studies.

Abstract

We report on the first seawater tests at 1 atm of the Honeywell Durafet® pH sensor, a commercially available ion sensitive field effect transistor (ISFET). Performance of this sensor was evaluated in a number of different situations including a temperature‐controlled calibration vessel, the MBARI test tank, shipboard underway mapping, and a surface mooring. Many of these tests included a secondary reference electrode in addition to the internal reference supplied with the stock Durafet sensor. We present a theoretical overview of sensor response using both types of reference electrode. The Durafet sensor operates with a short term precision of ± 0.0005 pH over periods of several hours and exhibits stability of better than 0.005 pH over periods of weeks to months. Our tests indicate that the Durafet pH sensor operates at a level of performance satisfactory for many types of biogeochemical studies at low pressure.

References

19