ASAIO Journal · 2002 · 77 citations · 12 references
Life SciencesCardiologyPhysiologyClinical CardiologyExercise PhysiologyPhysiological Systems ModelingCardiovascular System ModelingCirculatory PhysiologyHeart FailureHemodynamic SupportHemodynamicsHemodynamics ModelingCirculatory DisordersCirculatory StabilityCirculatory System ModelingExercise Circulatory ResponsesExercise HemodynamicsLeft Ventricular Assist DevicesNonlinear Dynamic ResponseVentricular Assist Device Support
This article presents an integrated model of the human circulatory system that incorporates circulatory support by a brushless DC axial flow ventricular assist device (VAD), and a feedback VAD controller designed to maintain physiologically sufficient perfusion. The developed integrated model combines a network type model of the circulatory system with a nonlinear dynamic model of the brushless DC pump We show that maintaining a reference differential pressure between the left ventricle and aorta leads to adequate perfusion for different pathologic cases, ranging from normal heart to left heart asystole, and widely varying physical activity scenarios from rest to exercise.
12
Textbook of Medical Physiology
Annals of Internal Medicine · 1966 · 9.8K citations
Life Sciences, Anatomy, Physiology +12
PHYSBE ... a physiological simulation benchmark experiment
J. McLeod · SIMULATION · 1966 · 78 citations
Modeling and identification of an axial flow blood pump
Seongjin Choi, J.R. Boston, D. Thomas et al. · 1997 · 66 citations