Medical Physics · 1994 · 114 citations · 0 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDetector TechnologyImagingMedical ImagingMedical Imaging PhysicsPositron Emission TomographyPositron Emission Tomography PhysicsDetector Spatial ResolutionFull Width At Half MaximumMagnetic Field InhomogeneityPet Detector TechnologyPositron Emission Tomography DetectorsPositron Emission Tomography InstrumentationSpatial Resolution MeasurementStrong-field Regime
Detector geometry, spatial sampling, and more fundamentally, positron range and noncollinearity of annihilation photon emission define Positron Emission Tomography (PET) spatial resolution. In this paper, a strong magnetic field is used to constrain positron travel transverse to the field. Measurement of the spread function from a 500 microns diameter 68Ga impregnated resin bead shows a squeezing of the full width at half maximum (FWHM) by a factor of 1.0, 1.22, 1.42, and 2.05, at 0, 4.0, 5.0, and 9.4 Tesla, respectively. The full width at tenth maximum (FWTM) decreases by a factor of 1.0, 1.73, 2.09, and 3.20, at 0, 4.0, 5.0, and 9.0 Tesla, respectively. Acquiring a PET image in a magnetic field should significantly reduce resolution loss due to positron range.