Journal of Orthopaedic Trauma · 1990 · 183 citations · 0 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryElbow SurgeryOrthopedicsShoulder SurgeryDistal Humerus Fracture SurgeryFixation BiomechanicsFracture FixationFracture StabilizationImplant FixationInternal FixationArticular ReconstructionBiomechanics Of Internal FixationDistal Humerus Fracture ReconstructionDual PlatingInternal Fixation DevicesPlate-screw ConstructsSingle Plate FixationSupracondylar Fractures
Three commonly used configurations of various implants used for fixation of distal humeral fractures were quantitatively compared. The double plate construct, irrespective of plate type (1/3 tubular and/or 3.5 mm reconstruction plate), was significantly stronger, both in rigidity and fatigue testing, than cross screws or the single "Y" plate. If rigid stabilization of supracondylar or bicondylar distal humeral fractures is desired, then two plate constructs, at right angles (the ulna plate medially, the lateral plate posteriorly), are biomechanically optimal.