Natural SciencesChemistryToxicologyAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisContaminant AnalysisDermal ToxicologyForensic AnalysisInstrumental AnalysisIonization TechniquesMass SpectrometryElectrospray IonizationExplosivesExplosives AnalysisExplosives ChemistryExplosives DetectionMass Spectrometric IdentificationTandem Mass SpectrometryTnt
Single nanogram amounts of the explosives TNT, RDX, HMX, PETN and their mixtures were detected and identified in a few seconds on the surface of human skin without any sample preparation by desorption electrospray ionization (DESI) using a spray solution of methanol-water doped with sodium chloride to form the chloride adducts with RDX, HMX, and PETN while TNT was examined as the radical anion and tandem mass spectrometry was used to confirm the identifications.