Minimally Invasive Application of the Non-Contact-Bridging (NCB) Plate to the Proximal Humerus: An Anatomical Study

48

Citations

18

References

2007

Year

Abstract

The minimal anterolateral acromial approach is suitable for MI technique to apply the NCB-PH. The relationship of the axillary nerve to the plate is anatomically close. We recommend that strict bone contact be maintained during plate insertion and that screw insertion complies with the guidelines provided for this technique. In a small clinical cases series, the plate and screws were inserted in accordance with these guidelines and no axillary nerve lesions have yet been detected.