Retrieval of cardiac phase from IVUS sequences
36
Citations
6
References
2003
Year
Cardiac MuscleIvus ImagesHeart FailureEngineeringMedical UltrasoundBiomedical EngineeringCardiac PhaseCardiologyBlood Flow MeasurementCardiac MechanicRadiologyHealth SciencesCardiovascular ImagingVascular ImageMedical ImagingIvus ImageMedical Image ComputingDigital Subtraction AngiographyPhysiologyBiomedical ImagingElectrophysiologyCoronary Artery
Most of the quantitative measures from Intravascular Ultrasound (IVUS) images vary with the cardiac cycle. Although ECG-gated acquisition can prevent the pulsations from influencing the measurements, it may extend the acquisition time, and furthermore, very few IVUS systems currently in clinical use incorporate ECG-gated function. In this paper, we present a practical method to retrieve cardiac phase information directly from in vivo clinical IVUS image sequences. In an IVUS image that contains a cross-section of coronary artery, there are three regions annularly distributed from the center of the image - catheter, lumen, and part of the vessel wall. The catheter region exhibits virtually no change from frame to frame during the catheter pullback. While the lumen is a dark region, the vessel wall region appears bright. The change in lumen size and position that accompanies the pulse causes the image intensity of the IVUS images to exhibit a periodic variation along the pullback path. By extracting this signal attributed to the cardiac cycle, a subsequence of frames during pullback at the same phase of the cardiac cycle can be selected. The method was tested by the IVUS images of both a coronary phantom and a patient.