Journal of Computer Assisted Tomography · 2009 · 19 citations · 13 references
Veterinary MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation MeasurementComputed TomographyComputed Tomography PhysicsMedical ImagingPhantom DosimetryRadiation Imaging SystemsRadiographic ImagingAnteroposterior ProjectionsComputed Tomography Dose MeasurementComputed Tomography Phantom DosimetryImage Noise AssessmentRadiation Dose EstimationRadiation Dose ManagementRadiation Dose ReductionTube Current Modulation
We describe the effect of the off-center position, the bowtie filter type, and the direction of scan projection radiograph on the radiation dose and image noise of 64-detector computed tomographic scans using the automatic tube current modulation technique. On anteroposterior scan projection radiographs, the tube current ratio increased, with an increase in the off-center position of the phantom irrespective of the filter size. On lateral scan projection radiographs, it remained constant with each bowtie filter type regardless of the off-center phantom position.
13
The influence of patient centering on CT dose and image noise