Image Noise and Radiation Dose Using an Automatic Tube Current Modulation Technique at 64-Detector Computed Tomography

Yoshinori Funama, Katsuyuki Taguchi, Kazuo Awai, Daisuke Sakabe, Masamichi Shimamura, Yasuyuki Yamashita

Journal of Computer Assisted Tomography · 2009 · 19 citations · 13 references

Concepts

Abstract

We describe the effect of the off-center position, the bowtie filter type, and the direction of scan projection radiograph on the radiation dose and image noise of 64-detector computed tomographic scans using the automatic tube current modulation technique. On anteroposterior scan projection radiographs, the tube current ratio increased, with an increase in the off-center position of the phantom irrespective of the filter size. On lateral scan projection radiographs, it remained constant with each bowtie filter type regardless of the off-center phantom position.