Journal of Computer Assisted Tomography · 2000 · 110 citations · 11 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage OptimizationImage Quality AssessmentImplant-site ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingArtifact ReductionArtifacts In Magnetic Resonance ImagingGradient Echo SequencesImaging Artifact AssessmentMagnetic Susceptibility MeasurementMetal Artifact CorrectionMetal Artifact ReductionSusceptibility Effects
PURPOSE: The purpose of this work was to determine the optimal imaging parameters for minimization of metallic susceptibility artifacts during gradient echo (GRE) imaging. METHOD: We performed GRE imaging of titanium screws in a nickel-doped agarose gel phantom, systematically varying several parameters to characterize and quantify susceptibility artifacts. RESULTS: The greatest reduction in artifact size came from using a short TE; increasing the frequency matrix and decreasing the slice thickness also contributed substantially to reducing the artifact size. Whenever possible, implanted prostheses should be aligned with the main magnetic field to minimize artifact size. Parameters with negligible effect on artifact size included bandwidth, phase encode matrix, and field of view. CONCLUSION: Radiologists can easily adjust the above parameters in their imaging protocols to improve GRE image quality in patients with implanted metallic devices.
11
MR imaging in patients with metallic implants.
R W Laakman, Benjamin Kaufman, Jisu Han et al. · Radiology · 1985 · 123 citations
Biopsy needle susceptibility artifacts
Mark E. Ladd, Peter Erhart, Jörg F. Debatin et al. · Magnetic Resonance in Medicine · 1996 · 121 citations