NONINVASIVE PULSE OXIMETRY IN CHILDREN WITH CYANOTIC CONGENITAL HEART DISEASE

Robert A. Boxer, ILENE BURSON GOTTESFELD, Sharanjeet Singh, Michael A. LaCorte, Vincent Parnell, Peter J. Walker

Critical Care Medicine · 1987 · 49 citations · 0 references

Concepts

Abstract

Arterial oxygen saturation, determined noninvasively by pulse oximetry in 32 pediatric patients with cyanotic congenital heart disease (CHD), was compared with oxygen saturation measured by a cooximeter in simultaneously obtained arterial blood samples. The patients were studied in the cardiac catheterization laboratory, operating room, and ICU. Excellent correlation by linear regression (n = 108, r = .95) was observed between the two methods at oxygen saturations ranging from 35% to 95%. These observations show that in infants and children with cyanotic CHD, arterial oxygen saturations can be determined accurately and reliably by pulse oximetry at rest and during changing circulatory states.