Neurocase · 2012 · 14 citations · 32 references
Life SciencesNeurologyNeuroscienceClinical NeuroscienceNeural EngineeringNeurophysiologyDeep Brain StimulationNeural Circuit ModelingNeural StimulationNeuromodulationNeurostimulationBipolar StimulationDeep Brain Stimulation DevicesElectric Field EffectsElectrical Field ModelingEssential TremorMulti-disciplinary TeamworkStimulation Parameter ProgrammingStimulation Parameters
The quadripolar electrodes used for deep brain stimulation are designed to give flexibility in contact configuration, optimize therapeutic effect, and minimize side-effects. A patient with essential tremor did not tolerate a bipolar setting due to the emergence of a pulling sensation in her face. However, when the polarity of the contacts was reversed, a 70% higher voltage was tolerated. Using an electric field model, we predicted that this effect was due to the proximity of the topmost contact to the internal capsule. Post-operative imaging supported this prediction. These results demonstrate how a multi-disciplinary approach allows us to optimize parameter settings.
32
Deep Brain Stimulation for Treatment-Resistant Depression
Helen S. Mayberg, Andrés M. Lozano, Valerie Voon et al. · Neuron · 2005 · 4.1K citations · Full text