Abstract

It is important to use a calibrated reference KAP-meter and a harmonised calibration method in the P(KA) calibration in hospitals. The obtained uncertainty in the P(KA) readings is comparable with other calibration methods if the information in the calibration certificate is correct used, corrections are made and proper positioning of the KAP-chamber is performed. This will ensure a reliable estimate of the patient dose and a proper optimisation of conventional x-ray examinations and interventional procedures.