Publication | Closed Access
Multi‐institutional dosimetric and geometric commissioning of image‐guided small animal irradiators
53
Citations
29
References
2014
Year
With the exception of the half-value layer, and mechanical properties, there were significant differences between the dosimetric and geometric properties of the three systems. This underscores the need for careful commissioning of each individual system for use in radiobiological experiments.