Publication | Closed Access
Da Vinci Robot Error and Failure Rates: Single Institution Experience on a Single Three-Arm Robot Unit of More than 700 Consecutive Robot-Assisted Laparoscopic Radical Prostatectomies
124
Citations
6
References
2007
Year
For all the advanced features the da Vinci system offers, it is surprisingly reliable. Throughout our RLRP experience, device failure resulted in case conversion, procedure abortion, and surgeon handicap in 0, 0.5%, and 0.4% of procedures, respectively. As such, a lowered device FR of 0.5% should be used when counseling patients undergoing RLRP. To avoid futile general anesthesia, a policy should be enforced to ensure that the da Vinci system is completely set up before the patient enters the operating room.
| Year | Citations | |
|---|---|---|
Page 1