IEEE Transactions on Nuclear Science · 1977 · 10 citations · 5 references
EngineeringNuclear EngineeringInstrumentationNuclear PhysicsRadiation DetectionDetector TechnologyNuclear DecayParticle DetectorsRadiation DetectorsScintillation DetectorsDecay TimeDecay Time AnalysisPulse Height SpectraPulse-shape DiscriminationSodium Iodide DetectorsTemperature DependenceTemperature ExcursionsTime Constant
Pulse shape discrimination is used in important NaI(Tl) detector applications, such as, Phoswich applications, particle separation, and pile-up suppression. These applications utilizing only pulse height spectra gated by a PSD window can be compromised by temperature excursions in the detector. We present measurements of the NaI(Tl) decay time constant over the range 0°C - 24°C. A temperature variation of ~5 nsec/°C was measured. The implications of the temperature variation on detector performance are discussed.
5