Review of Scientific Instruments · 2013 · 22 citations · 20 references
EngineeringElectrical EngineeringMechanical EngineeringAnalog ElectronicsElectromagnetismInstrumentationAnalog InstrumentationInstrumentation AmplifiersLow-noise Circuit DesignNoise In Electronic DevicesSeismic InstrumentationSensor NoiseAccelerometersAccelerometryOperational Amplifier CircuitsOperational Amplifier DesignOperational Transconductance Amplifier DesignSeismic Noise AnalysisTranslation NoiseVoltage
A simple high-voltage circuit with a voltage range of 0 to 900 V and an open-loop bandwidth of 11 kHz is realized by using an operational amplifier and a MOSFET combination. The circuit is used for the levitation of a test mass of 71 g, suspended below the top-electrodes with a gap distance of 57 μm, so that the performance of an electrostatic accelerometer can be tested on the ground. The translation noise of the accelerometer, limited by seismic noise, is about 4 × 10(-8) m/s(2)/Hz(1/2) at 0.1 Hz, while the high-voltage coupling noise is one-order of magnitude lower.
20
Tests of the Gravitational Inverse-Square Law below the Dark-Energy Length Scale
D. J. Kapner, T. S. Cook, E. G. Adelberger et al. · Physical Review Letters · 2007 · 983 citations · Full text
Natural Sciences, Astronomy, Physics +15
Disturbance reduction requirements for LISA
Bonny L. Schumaker · Classical and Quantum Gravity · 2003 · 131 citations