Dynamic tomographic digital subtraction angiography using temporal filtration.

Robert A. Kruger, James A. Nelson, D Ghosh-Roy, F J Miller, R E Anderson, P Y Liu

Radiology · 1983 · 25 citations · 0 references

Concepts

Abstract

A variation of temporal recursive filtering for intravenous and intra-arterial digital subtraction angiography (DSA) is described. It results in a dynamic tomographic DSA display that may improve intravenous DSA performance by removing the confusion introduced by vessel overlap and reducing the severity of motion artifacts. A conventional circular tomographic apparatus was modified by the addition of a 23-cm image intensifier-television chain and made to run continuously at one revolution each 1.5 seconds. The video output was processed by the recursive filter in a two-step process resulting in a dynamic tomographic display of arterial opacification. Theory and results of initial animal studies are presented.