Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 1992 · 50 citations · 9 references
Bone DiseaseCementationCement—bone Interface StrengthPmma CementOsteoarthritisBone CementBone RemodelingSurgeryOsteoporosisShear StrengthBone DensityMedicineOrthopaedic Surgery
Interfacial shear strength between poly(methyl methacrylate) (PMMA) bone cement and cancellous bone was measured in bone samples from human proximal femora. Samples were prepared with fresh cement-bone, fresh cement inside a mantle of existing cement and with fresh cement-revised bone surfaces. Push-out tests to measure shear strength caused failure only at bone-cement interfaces; revised bone interfaces were 30 per cent weaker (P < 0.02) than primary interfaces. The clinical relevance is that revision of cemented joint arthroplasties may necessitate removal of components with sound cement-bone fixation. The practice of removing all traces of PMMA cement may not yield the optimal fixation; adhesion of fresh cement to freshly prepared surfaces of the existing cement might also be considered where circumstances are favourable.