Revision of Cemented Fixation and Cement—Bone Interface Strength

Alexander D. Rosenstein, Warren Macdonald, Agis A. Iliadis, P. McLardy-Smith

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 1992 · 50 citations · 9 references

Concepts

Abstract

Interfacial shear strength between poly(methyl methacrylate) (PMMA) bone cement and cancellous bone was measured in bone samples from human proximal femora. Samples were prepared with fresh cement-bone, fresh cement inside a mantle of existing cement and with fresh cement-revised bone surfaces. Push-out tests to measure shear strength caused failure only at bone-cement interfaces; revised bone interfaces were 30 per cent weaker (P < 0.02) than primary interfaces. The clinical relevance is that revision of cemented joint arthroplasties may necessitate removal of components with sound cement-bone fixation. The practice of removing all traces of PMMA cement may not yield the optimal fixation; adhesion of fresh cement to freshly prepared surfaces of the existing cement might also be considered where circumstances are favourable.