Abstract

An optical fibre sensing system based on a hybrid cascaded long period grating (LPG) and fibre Bragg grating configuration and a thermo-stabilized flow cell for refractometric measurements is proposed. The system makes it possible to measure, and thus to cancel, the LPG cross-sensitivities to strain, temperature and fibre bending. The experimental results show that the proposed system provides satisfactory performances as far as the refractive index sensitivity and resolution are concerned. The maximum sensor sensitivity and resolution are 3120 nm/RIU and 2 × 10 -5 RIU, respectively. The whole system with its flow cell and the gratings fabrication are extensively described, together with the acquisition and data processing. The stability of the sensor for several hours was also tested. We believe that the proposed system can be successfully used for label-free chemical/biochemical sensing.