Concepedia

Publication | Closed Access

Application of <scp>L</scp>‐(+)‐Lactate electrode for clinical analysis and monitoring of tissue culture medium

25

Citations

9

References

1985

Year

Abstract

L-(+)-Lactate oxidase (EC 1.1.3.2) was immobilized onto the porous side of a cellulose acetate membrane with asymmetric structure which has selective permeability to hydrogen peroxide. The lactate electrode was constructed by combination of a hydrogen peroxide electrode with the immobilized enzyme membrane. Properties of the enzyme membrane and characteristics of the lactate electrode were clarified for the determination of L-(+)-lactic acid. The lactate electrode responded linearly to L-(+)-lactic acid over the final concentration 0-0.25 mmol/L within 30 s. When the enzyme electrode was applied to the determination of L-(+)-lactic acid in control serum, within-day precision (CV), analytical recovery, and correlation coefficient between the electrode method and the colorimetric method were 1.4% with a mean value of 4.54 mmol/L, 98.0%, and 0.986, respectively. The lactate electrode was sufficiently stable to perform 1040 assays over 13 days operation for the determination of L-(+)-lactic acid. The dried immobilized enzyme membrane retained 84% of its initial activity after storage at 4 degrees C for 12 months. Moreover, the enzyme electrode was applied to the monitoring of culture medium for human melanoma cells. L-(+)-Lactate production and D-glucose consumption were closely related to cell numbers.

References

YearCitations

Page 1