American Journal of Roentgenology · 2007 · 101 citations · 11 references
MedicineRadiation OncologyRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryComputed TomographyComputed Tomography PhysicsMedical ImagingPatient Dose OptimizationPhantom DosimetryRadiation Dose MeasurementImage Noise AssessmentPatient Dose MeasurementPatient Dose ReductionPatient Radiation Dose OptimizationPhantom-based Dose MeasurementRadiation Dose EstimationRadiation Dose OptimizationTube Current Modulation
OBJECTIVE: Our objective was to develop a better understanding of the complex interrelationship between image noise, reconstruction slice thickness, and patient radiation dose on a 64-MDCT scanner that uses automated tube current modulation. MATERIALS AND METHODS: We reviewed physics theory and performed phantom dose measurements on a 64-MDCT scanner while altering operator-selectable image noise and reconstruction slice thickness. RESULTS: Using phantom dose measurements to adjust theoretic predictions, we constructed both a spreadsheet and a graph that visually display the interrelationships between operator-selected image noise and reconstruction slice thickness and the resulting patient dose. CONCLUSION: This table and graph may help operators understand the trade-offs when prospectively trying to minimize dose and optimize image noise for selected reconstruction slice thicknesses on this type of 64-MDCT scanner.
11