Publication | Closed Access
Comparison of Tissue Doppler Velocities Obtained by Different Types of Echocardiography Systems: Are They Compatible?
21
Citations
17
References
2010
Year
Although diastolic TD velocities had excellent correlations between the two machines, there was a systematic overestimation by the Sonos system. Since the limits of agreement do not allow replacing the measurements, we suggest using the same echocardiographic equipment at patient follow-up.