Contrast Media & Molecular Imaging · 2006 · 45 citations · 34 references
Natural SciencesCell BiologyRadiologyBiomedical ImagingDiagnostic ImagingMagnetic Resonance BiophysicsContrast AgentsMagnetic Resonance Contrast MechanismsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMedical ImagingEcho Time DependenceGradient Echo SequencesIron Oxide NanoparticlesMagnetic Oxide NanoparticlesMagnetic Resonance Imaging DetectionNanoparticle-based Contrast AgentsSpin Echo ImagingSpin Echo Sequences
Recent studies have shown that cell migration can be monitored in vivo by magnetic resonance imaging after intracellular contrast agent incorporation. This is due to the dephasing effect on proton magnetization of the local magnetic field created by a labelled cell. Anionic iron oxide nanoparticles (AMNP) are among the most efficient and non-toxic contrast agents to be spontaneously taken up by a wide variety of cells. Here we measured the iron load and magnetization of HeLa tumour cells labelled with AMNP, as a function of the external magnetic field. High-resolution gradient echo 9.4 T MRI detected individual labelled cells, whereas spin echo sequences were poorly sensitive. We then conducted a systematic study in order to determine the gradient echo sequence parameters (echo time, cell magnetization and resolution) most suitable for in vivo identification of single cells.
34