Review of Scientific Instruments · 1992 · 12 citations · 1 references
EngineeringNuclear EngineeringPhysicsNuclear PhysicsRadiation DetectionRadiation MeasurementDetector TechnologyNuclear FusionRadiation Detection PrinciplesRadiation DetectorsScintillation DetectorsTokamak EngineeringAbsorbed PowerPower DensityRare-earth-doped ScintillatorsResponse TimeScintillationScintillation DetectionScintillation EfficiencyTokamak Physics
The light output of ZnS scintillators used to detect escaping fusion products in the TFTR Tokamak is studied with 3.5-MeV alpha and 3-MeV proton beams. The emitted light first increases linearly with beam current and then saturates. In all cases investigated, the saturations start at a fairly constant absorbed power density of about 1 mW/cm2. The scintillators have adequate time response up to 50–100 kHz.
1
Radial diffusion coefficient for counter-passing MeV ions in the TFTR tokamak
S. J. Zweben, R. L. Boivin, C. S. Chang et al. · Nuclear Fusion · 1991 · 44 citations