Medical Physics · 1984 · 27 citations · 0 references
MedicineNuclear EngineeringRadiation OncologyRadiation DosimetryRadiation MeasurementRadiation SafetyExternal Beam DosimetryNuclear Medicine PhysicsPatient-specific DosimetryPhoton DosimetryRadiation Dose MeasurementRadiation Risk AssessmentBeta CountingBeta DosimetryBeta Radiation DosimetryClinical Photon Beam DosimetryDose Equivalent MeasurementNeutron Equivalent DoseNeutron LeakageParticle Beam Dosimetry
The McCall method has been used to measure neutron leakage from the Mevatron 77, 18- and 15-MV photon beams. Gold foil activation has been used employing a beta counting technique for the 18-MV beam and a gamma counting technique for both the 18- and 15-MV beam. The two counting techniques were used to evaluate their relative merit. The measurements were made at various locations in the patient-treatment plane for different field sizes. The results show that the thermal-neutron dose equivalent contributes only about 1%-2% of the total neutron dose equivalent. At 100 cm, the neutron dose equivalent for the 18-MV beam is approximately six times that of the 15-MV beam, slightly exceeding the 0.1% of the useful beam criteria used by some of the regulatory agencies. In light of the uncertainty in fluence to dose equivalent conversion factors, the increased dose equivalent above 0.1% is insignificant.