Analytical Letters · 1994 · 38 citations · 8 references
Chemical EngineeringBiomedical SensorsEngineeringBiosensing SystemsBiomedical DiagnosticsAbstract Amperometric BiosensorsImmobilised PeroxidaseBioelectronicsBioelectrochemistryTtf-modified BiosensorsAnalytical ChemistryBiomedical EngineeringElectroanalytical SensorNanosensorWearable BiosensorsChemical SensorHydrogen PeroxideElectrochemistry
Abstract Amperometric biosensors for hydrogen peroxide were produced, based on TTF-modified graphite disc electrodes, incorporating immobilised peroxidase. At the optimum operating potential (−200 mV vs. calomel electrode), a linear response to hydrogen peroxide was obtained in the range 0–2 mM (Km value = 1.98 mM). Low sensitivity to pH fluctuations and good stability at room temperature were exhibited.
8
Biosensors: principles and practice.
Jeffrey D. Newman, Anthony Turner · PubMed · 1992 · 28 citations