A Comparison of Frontal and Occipital Bispectral Index Values Obtained During Neurosurgical Procedures

Toshie Shiraishi, Hiroyuki Uchino, Takeshi Sagara, Nagao Ishii

Anesthesia & Analgesia · 2004 · 62 citations · 5 references

Concepts

Abstract

UNLABELLED: We placed bispectral index (BIS) sensors on the frontal and occipital areas of neurosurgical patients and compared BIS values obtained from both areas during propofol/fentanyl anesthesia. BIS showed a strong correlation between frontal and occipital montages (r(2) = 0.96; P = 0.03). It may be valid to measure BIS with the sensor on the occipital area if required during frontal neurosurgical procedures. IMPLICATIONS: Bispectral values were positively correlated when recorded from frontal and occipital sensors in patients undergoing clipping of unruptured cerebral aneurysms while anesthetized with propofol and fentanyl.

References

5