Journal of the Optical Society of America · 1963 · 28 citations · 4 references
EngineeringMechanical EngineeringRadiation DetectionRadiation MeasurementSensorsCalorimetryDetector TechnologyRadiation DetectorsRadiation InstrumentationThermal SensorsCalorimetric DetectorsEmissivityEmissivity EstimationReflectanceSpectral Response CharacterizationThermal DetectorsThermocouple InstrumentationThermocouples
A calorimeter in the form of a 15° cone has been constructed and evaluated for use as a “black” radiation detector. The internal surface of the cone is coated with colloidal graphite to enhance the absorptance. Spectral emissivity and reflectance measurements of the blackening material are used to estimate the effective absorptance of the conical receiver. To reduce many of the difficulties associated with a dc system, this detector is used with an ac amplifying system at a frequency of 1 cps. The spectral response of three commercial thermocouples is compared with that of the conical receiver.
4