Annals of the New York Academy of Sciences · 1999 · 21 citations · 9 references
EngineeringBiomedical EngineeringNeuroscienceBiomedical ImagingNeural EngineeringNeurophysiologyBiopotential MeasurementBrain ImagingElectroencephalographyMedical Imaging InstrumentationNeural Signal AcquisitionBrain Function MappingElectrical ImpedanceElectrical Impedance TomographyElectroencephalography InstrumentationRecording ElectrodesSignals
An EIT system has been produced that has been optimized for imaging impedance changes with scalp electrodes during brain activity in ambulant subjects. It can record from 225 Hz to 65 kHz, has a small headbox on a lead 10 m long, and has software programmable electrode selection. In calibration experiments in a small cylindrical tank filled with potassium chloride solution and samples of cucumber, noise was less than 1% with averaging, and acceptable images were produced at frequencies down to 1800 Hz. This suggests that EIT can be performed at low frequencies, which are likely to give larger signals during brain activity. Future work will include trials in humans and improvement of the current source and isolation.
9
Skin impedance from 1 Hz to 1 MHz
J. Rosell, J. Colominas, Pere J. Riu et al. · IEEE Transactions on Biomedical Engineering · 1988 · 341 citations
Engineering, Biomedical Engineering, Electrical Engineering +10