Radiation dose and image information in computed radiography: A phantom study of angle measurements in the weight-bearing knee
14
Citations
10
References
2000
Year
No significant correlation was found between patient dose and measuring accuracy within the evaluated exposure interval. At the lowest exposure of the CR system, the energy imparted to the patient was 30 microJ. Compared with our conventional analogue method this is a reduction by 98%. The effective dose was as low as 1.5 microSv. The CR technique creates possibilities to adapt exposure parameters, and thus the radiation dose to the patient, according to the purpose of the investigation.