Review of Scientific Instruments · 2002 · 58 citations · 9 references
EngineeringNuclear EngineeringNuclear InstrumentationNuclear PhysicsRadiation DetectionDetector TechnologyGamma-ray DetectionNeutron DetectionRadiation DetectorsReactor InstrumentationScintillation DetectorsDetector ConstructionNeutron Detector SystemsNeutron DetectorsNeutron DiagnosticsNeutron Emission SpectraNeutron-gamma DiscriminationScintillator CrystalsTokamak Physics
This article describes a diagnostic for measuring neutron emission profile in JT-60U. The Stilbene neutron detector, developed by TRINITI laboratory in Russia, has been installed on the JT-60U Tokamak to measure the neutron emission profile for the first time. The Stilbene neutron detector is a detector which combines a Stilbene crystal scintillator with a neutron-gamma pulse shape discrimination circuit, with a very compact size. Performance tests were carried out using neutron and gamma-ray source prior to installation on JT-60U. Good gamma suppression of the Stilbene neutron detector was verified. Though the neutron emission profile obtained by Stilbene neutron detectors has error of 30% in innermost channel with a calculation using measured plasma parameters, there is an agreement within 10% error in the other channels.
9
Development of organic scintillators
F.D. Brooks · Nuclear Instruments and Methods · 1979 · 353 citations
Heating of toroidal plasmas by neutral injection
Thomas H. Stix · Plasma Physics · 1972 · 343 citations