Electrical Methods in Occlusal Caries Diagnosis: An in vitro Comparison with Visual Inspection and Bite–Wing Radiography

M.C.D.N.J.M. Huysmans, Ch. Longbottom, Nigel Pitts

Caries Research · 1998 · 62 citations · 12 references

Concepts

Abstract

The aim of this study was to compare in vitro diagnostic performance of three electrical methods for occlusal caries diagnosis with that of visual inspection and bite-wing radiography. One hundred and seven extracted molar and premolar teeth were subjected to the diagnostic methods by 2 operators and subsequently sectioned for histological validation. Electrical measurements were made at site level and at surface level using two different instruments. The diagnostic parameters calculated from the results were: sensitivity, specificity, diagnostic accuracy at a theoretical caries prevalence of 10% and area under the ROC curve. The electrical methods and bite-wing radiography showed higher sensitivity and lower specificity than visual inspection. Diagnostic accuracy was significantly lower for bite-wing radiography and one electrical method than for visual inspection. Overall diagnostic performance of two electrical methods was superior.