Journal of Analytical Atomic Spectrometry · 2002 · 81 citations · 23 references
Natural SciencesEarth SciencesEnvironmental SciencesAnalytical ChemistryElemental AnalysisGeochemistryEnvironmental Elemental AnalysisEnvironmental SamplingRadiochemical AnalysisSediment ChemistrySediment CompositionStable Isotope AnalysisEnvironmental IsotopesEnvironmental Sample AnalysisIsotopic AbundancesIsotopic AnalysisIsotopic Composition AnalysisLower Limit Of QuantificationPrecision MeasurementStandard Reference Materials
A method for the separation and purification of environmental levels of plutonium and neptunium from totally digested sediment samples and determination of isotopic abundances of 237Np, 239Pu, 240Pu and 241Pu by ICP-MS has been developed. Data from the analysis of standard reference materials (SRM-4350B, IAEA-135, and IAEA-384), selected sediments collected from the Ob River, Siberia, and laboratory intercomparison analyses conducted with scientists at DOE's Environmental Measurements Laboratory using TIMS are presented and discussed. Chemical recoveries for Pu and Np are 90 and 70 percent, respectively. Measurement precision ranged between 1 and 5 percent for all isotopes except 241Pu, for which it was 10–15 percent. The lower limit of detection was 0.5 fg g−1 for all isotopes of interest.
23
Yasuyuki Muramatsu, W. Rühm, Satoshi Yoshida et al. · Environmental Science & Technology · 2000 · 172 citations
The240Pu239Pu ratio, a potential geochronometer
Minoru Koide, K.K. Bertine, Tsaihwa J. Chow et al. · Earth and Planetary Science Letters · 1985 · 147 citations
J. Sabine Becker, Hans‐Joachim Dietze · Journal of Analytical Atomic Spectrometry · 1999 · 117 citations