Modulation Transfer Function Associated with Geometrical Unsharpness in Medical Radiography

G. Lubberts, Kurt Rossmann

Physics in Medicine and Biology · 1967 · 28 citations · 9 references

Concepts

Abstract

The geometrical unsharpness caused by the size of the focal spot and the geometry of the exposure set-up is expressed in terms of the modulation transfer function (MTF) by means of Fourier techniques. The theoretical MTF associated with geometrical unsharpness has been verified experimentally with light and a small aperture to simulate a uniform focal spot. The MTF associated with the unsharpness produced by an actual focal spot was measured by means of edge exposures.

References

9