Magnetic Resonance Imaging with Implanted Neurostimulators: An In Vitro and In Vivo Study

121

Citations

9

References

1999

Year

Abstract

MRI can be safely performed in patients with implanted neurostimulation systems with the tested deep brain leads connected to an IPG (ITREL II and 3), with running parameters. No heat induction was detected, and the experimentally measured induced voltage did not seem to harm the patients. Only the reed switch of the IPGs was activated; the other parameters remained unchanged. Further investigations must be performed to study the local electrical effects in larger plate electrodes; these effects might cause slight discomfort. There is no danger with any type of electrode during MRI examinations if the electrodes lie outside the region of interest. These observations are restricted to the tested devices. A conscientious estimation of the risks and benefits of MRI for patients with implanted devices is recommended. If the type of device is not known to the examiner, MRI should still be considered to be contraindicated.