Design and development of C-arm based cone-beam CT for image-guided interventions: initial results

Guang‐Hong Chen, Joseph Zambelli, Brian Nett, Mark Supanich, Cyril Riddell, Barry Belanger, Charles A. Mistretta

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2006 · 30 citations · 0 references

Concepts

Abstract

X-ray cone-beam computed tomography (CBCT) is of importance in image-guided intervention (IGI) and image-guided radiation therapy (IGRT). In this paper, we present a cone-beam CT data acquisition system using a GE INNOVA 4100 (GE Healthcare Technologies, Waukesha, Wisconsin) clinical system. This new cone-beam data acquisition mode was developed for research purposes without interfering with any clinical function of the system. It provides us a basic imaging pipeline for more advanced cone-beam data acquisition methods. It also provides us a platform to study and overcome the limiting factors such as cone-beam artifacts and limiting low contrast resolution in current C-arm based cone-beam CT systems. A geometrical calibration method was developed to experimentally determine parameters of the scanning geometry to correct the image reconstruction for geometric non-idealities. Extensive phantom studies and some small animal studies have been conducted to evaluate the performance of our cone-beam CT data acquisition system.