Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2006 · 30 citations · 0 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImage-guided RadiotherapyImage-guided TherapyRadiation Imaging SystemsRadiographic ImagingComputed Tomography AcquisitionComputed Tomography CalibrationCone-beam Computed Tomography (Cbct)Cone-beam Computed Tomography ReconstructionContrast ResolutionCt Image ReconstructionSmall Animal ImagingX-ray Imaging Systems
X-ray cone-beam computed tomography (CBCT) is of importance in image-guided intervention (IGI) and image-guided radiation therapy (IGRT). In this paper, we present a cone-beam CT data acquisition system using a GE INNOVA 4100 (GE Healthcare Technologies, Waukesha, Wisconsin) clinical system. This new cone-beam data acquisition mode was developed for research purposes without interfering with any clinical function of the system. It provides us a basic imaging pipeline for more advanced cone-beam data acquisition methods. It also provides us a platform to study and overcome the limiting factors such as cone-beam artifacts and limiting low contrast resolution in current C-arm based cone-beam CT systems. A geometrical calibration method was developed to experimentally determine parameters of the scanning geometry to correct the image reconstruction for geometric non-idealities. Extensive phantom studies and some small animal studies have been conducted to evaluate the performance of our cone-beam CT data acquisition system.