Abstract

Cancellous bone density and total cancellous screw depth penetration seem to be critical variables. Although the patients' bone density cannot be controlled, surgeons may decrease the risk of failure by maximizing the length of the screws within the cancellous bone. Analysis of microcrack formation revealed that failures begin at the midportion and tips of the inferior screws and at the bone-plate interface of the inferior screws.