Chopper system for time resolved experiments with synchrotron radiation

Marco Cammarata, Laurent Eybert, Friederike Ewald, Wolfgang Reichenbach, Michaël Wulff, Philip Anfinrud, Friedrich Schotte, Anton Plech, Qingyu Kong, Maciej Lorenc,

Review of Scientific Instruments · 2009 · 145 citations · 17 references

Concepts

Abstract

A chopper system for time resolved pump-probe experiments with x-ray beams from a synchrotron is described. The system has three parts: a water-cooled heatload chopper, a high-speed chopper, and a millisecond shutter. The chopper system, which is installed in beamline ID09B at the European Synchrotron Radiation Facility, provides short x-ray pulses for pump-probe experiments with ultrafast lasers. The chopper system can produce x-ray pulses as short as 200 ns in a continuous beam and repeat at frequencies from 0 to 3 kHz. For bunch filling patterns of the synchrotron with pulse separations greater than 100 ns, the high-speed chopper can isolate single 100 ps x-ray pulses that are used for the highest time resolution. A new rotor in the high-speed chopper is presented with a single pulse (100 ps) and long pulse (10 micros) option. In white beam experiments, the heatload of the (noncooled) high-speed chopper is lowered by a heatload chopper, which absorbs 95% of the incoming power without affecting the pulses selected by the high speed chopper.

References

17