Accuracy evaluation of surface-based registration methods in a computer navigation system for hip surgery performed through a posterolateral approach

84

Citations

18

References

2001

Year

Abstract

Slice thickness and reconstruction pitch affected the accuracy of registration. As the sampling area was expanded from the periarticular area to the distant peripheral area, accuracy increased slightly. Accuracy did not increase when the whole area was used, but in fact decreased, especially in the femur. The positive effect of increasing the number of sampling points was saturated at 30 points when the surface of the periarticular area was sampled. The following trade-off between accuracy and invasiveness, in terms of various parameters of preoperative and intraoperative data, is proposed as clinically optimal: perform the CT scan with 3-mm slice thickness and 1-mm reconstruction pitch, and sample the periarticular area with 30 sampling points. With these parameters, the accuracy of registration was 1.2 mm and 0.9 degrees of bias with 0.7 mm and 0.3 degrees of RMS in the pelvis, and 1.4 mm and 0.6 degrees of bias with 1.3 mm and 0.3 degrees of RMS in the femur.