2014 · 10 citations · 8 references
EngineeringBioengineeringBiomedical EngineeringBioinstrumentationBiomemsNeural EngineeringBioelectronic MedicineBiomedical InstrumentationElectromyogram AnalysisMuscle Signal ProcessingNeural Signal AcquisitionNeural StimulationBiofeedback In DysphagiaDuty Cycle ControlElectrical Muscle StimulationElectrical StimulationElectrical Stimulation ParametersElectromyography InputElectromyography InstrumentationStimulation Parameters
A dedicated front-end for biofeedback neuromuscular electrical stimulation (NESM) system is proposed. For controllable dysphagia treatment, the integrated circuit (IC) provides a stimulator front-end with programmable stimulation parameters (2µA–1mA current amplitude, DC-2KHz frequency, and variable duty cycle) and an electromyogram/impedance (EMG/ETI) readout front-end with programmable gain and bandwidth (42–80dB, 0.1Hz–1.2KHz) for biofeedback. Area-efficient, low-power but high precision current controlling is achieved by inducing the sigma-delta modulator technique in the stimulation channel. The measured impedance and EMG signal are used to determine the stimulation parameters, enabling a closed loop optimized treatment. The proposed front-end is fabricated in a 0.18 µm standard CMOS process technology and dissipates a peak power of 2.3 mW at the supply voltage of 1.8 V. Measurement results on a live person are also provided to validate the system's effectiveness.
8
Sensory Transcutaneous Electrical Stimulation Improves Post-Stroke Dysphagic Patients
Syrine Gallas, Jean Paul Marie, Anne‐Marie Leroi et al. · Dysphagia · 2009 · 128 citations