Virtual CT Intravascular Endoscopy of the Aorta: Pierced Surface and Floating Shape Thresholding Artifacts
18
Citations
12
References
1999
Year
Computed TomographyShape ArtifactsVascular ImageHealth SciencesMedical ImagingEndovascular TechniqueMedicineMean AttenuationVirtual RealityVascular SurgeryCt ScanInterventional RadiologySurgeryThreshold LevelsDigital Subtraction AngiographyRadiologyCardiovascular Imaging
Two types of artifacts may appear in virtual computed tomographic endoscopic views of the aorta rendered at different threshold levels: pierced surface and floating shape artifacts. A positive correlation was found between mean attenuation of the aorta and the threshold levels at which these artifacts appeared. The correlation was statistically significant (0.71 < or = r < or = 0.86) for floating shape. An artifact-free threshold range can be predicted on the basis of aortic enhancement.